잠시만 기다려 주세요. 로딩중입니다.

혈청 Triglyceride 측정에 관한연구 -효소법과 Acetylacetone 법과의 비교-

A Study on Determination of Triglyceride. -Comparision of enzymatic and acetylacetone method-

대한병리학회지 1979년 13권 3호 p.199 ~ 202
김상호, 조한의,
소속 상세정보
김상호 (  ) - 충남대학교
조한의 (  ) - 서울대학교 의과대학 임상병리과

Abstract


A comparision of analytical methods for the determination of serum triglyceride was
made between enzymatic and acetylacetone methods.
The authors evaluated intra- and inter-assay variation as well as correlation between
two methods.
The results are as follows.
1. The mean values of triglyceride of one commercial control serum (Monitrol I) were
98. 02mg/dl by enzymatic method and 68.54mg/dl by acetylacetone method: enzymatic
method shown higher value compared to acetylacetone method by 30mg/dl.
2. Intra and inter-assay variation (n=12) represented by coefficient of variation was 3.1%
and 5.05% by enzymatic method, and 7.7% and 15.86% by acetylacetone method;
enzymatic method shorts lower inter-and inter assay variation.
3. A correlation between two methods (n=30) was made: correlation coefficient r=0.982,
and linear regression y= 1.22x+9.32