炭水素元素分析에 있어서 檢體充塡位置에 基因하는 誤差에 對한 硏究

Studies on the Errors, in the Microdetermination of Carbon and Hydrogen, due to the Positions of Sample

약학회지 1962년 6권 2호 p.23 ~ 28
임중기,
소속 상세정보
임중기 (  ) - Sung Kyun Kwan University College of Pharmacy

Abstract


To reduce the errors, in the microdetermination of carton and hydrogen, by the electrical automatic combustion apparatus, measured the errors due to the positions of sample in the carbon and hydrogen combustion tute.
The samples used are glucose, acetanilide, phenacetin and diazoaminobenzene. Experimented positions of samples were 3㎝, 5㎝, 7㎝, 8㎝, and 9㎝ from the front edge of the long stationary burner.
In this experinent, it was found the the most suitable position of sample in the carter and hydrogen combustion tube is 8㎝ from the long stationary burner.
(Received September 15, 1962)