迅速 Cephalin-Cholesterol Flocculation 檢査法에 대한 考察

A Comparative study on a rapid Cephalin-cholesterol Flocculation Test

중앙의학 1968년 14권 2호 p.129 ~ 131
문재은,
소속 상세정보
문재은 (  ) - 가톨릭의과대학 임상병리학교실

Abstract


In 111 in and out-patients at the St. Mary´s Hospital, Catholic Medical College, Seoul the Hanger´s original cephalin-cholesterol flocculation test was compared with the new Rosenberg´s rapid method employing four hours refrigeration and a spectrophotometricc reading.
Under our Inboratory conditions, the Rosenberg´s method revealed to be slightly less sensitive than that of Hanger. However, in the range considered abnormal by the Hanger´s, method, the results were well correlated with those by the Rosenberg´s, and the mode of correlation was steady and constant. Moreover, the rapid method required a considerably shorter time, eliminating the possible error in the Hanger´s method due to visual reading, and therefore was considered to be of value as a liver function test in the clinical laboratory.